Skin / Manual
(MODEL: IRW – 1803D)

Skin / Air / Manual  (LCD)
(MODEL: IRW – 2809D)

  • 1” bright skin temp. display
  • Set temp. display
  • % of heater output display on Bar Graph
  • Selection of Skin & Manual modes
  • Temperature display in °C or °F
  • APGAR Timer
  • Feed timer
  • Audio Alarm mute facility
  • Mute time programmable

Manual Mode

  • Time programmable
  • Heater output adjustable
  • Over temp. cutoff.

Skin Mode
Safety alarms
(with status LED)
Temp. High (>1°C of set temp.)
Temp. Low (<1°C of set temp.)
Temp. Over Range (>38°C)
Temp. Under Range (<34°C)
(Actual temp. starts blinking)
Probe Fail
Power failure
Heater Failure
System Failure

Safety high temp. cutoff at 38°C

  • 1” bright skin temp. display
  • 1” bright air temp. display
  • % of heater output display on Bar Graph
  • Temperature display in °C or °F
  • Set temp. display on LCD
  • Control mode display on LCD
  • APGAR Timer display on LCD
  • Feed Timer
  • Baby attending alarm
  • Mute time programmable
  • Audio Alarm mute facility

Manual Mode

  • Time programmable
  • Heater output adjustable
  • Over temp. cutoff.

Safety alarms, LCD messages

Skin Mode
Skin Temp. High (>1.5°C of set temp.)
Skin Temp. Low (<1.5°C of set temp.)
Skin Temp. Over Range (>38°C)
Skin Temp. Under Range (<34°C)
Skin Probe Fail

Air Mode
Air Temp. High (>1.5°C of set temp.)
Air Temp. Low (<3.0°C of set temp.)
Air Temp. Over Range (>39°C)
Air Probe Fail

Safety high temp. cutoff at

  • Skin Temp.> 38°C
  • Air Temp.> 39°C
  • Power Fail
  • Heater Fail
  • System Fail